I. Equipment maintenance
1.Power off the equipment and open the front cover.
2.Ensure the equipment is powered off and cooled down before maintenance.
3.Remove the water level plate and clean it. (Models with a chamber volume ≥110L are standard-equipped with a water level plate removal tool. One end of the tool is for taking out the water level plate, and the other end fitted with a silicone brush facilitates cleaning of the chamber, water level sensor and heating tube.)
4.Cleaning of sterilization chamber: Clean the sterilization chamber regularly to remove limescale and debris. Use a handled brush to clean the bottom of the chamber. Do not apply excessive force to avoid damaging the electric heating tube and temperature control switch.
Recommendation: Clean once a week.
5.Cleaning of water level sensor: Keep the water level sensor clean. Dirt attached to the sensor surface may cause false alarms or even equipment shutdown.
Recommendation: Wipe the sensor surface with a soft cloth once a week to remove contaminants.
6.Cleaning of electric heating tube
Recommendation: Clean once a month. Check the surface of the electric heating tube. If dirt is present, gently scrub with a soft brush or scouring pad. Take care not to shift or damage the temperature control switch during scrubbing.
ZEALWAY provides quality autoclaves, full services and professional sterilization solutions for you.