During the operation of the autoclave, operators frequently encounter and urgently need to resolve error codes. Rapid and accurate fault diagnosis and timely troubleshooting are essential for instrument operators.
- E01
Possible Cause: The chamber door is not locked.
Solutions:
1.GWA/GF Series: Check whether the chamber door is tightly closed and engaged.
2.GR Series: Verify that the handle is fully pulled into position.
3.GI Series: Confirm that the handle is fully rotated into position.
- E04
Possible Cause: The chamber door is not locked.
Solutions:
1.GWA/GF Series: Check if the chamber door is tightly pressed and engaged.
2.GR Series: Check whether the handle is pulled fully into position.
3.GI Series: Verify that the handle is rotated to the locking position.
- E10
Possible Cause: Abnormal events such as power failure occur during sterilization.
Solutions: Shortpress the “STOP” key to clear the error code, open and close the door again, then restart the sterilization cycle.
- E15
Possible Causes:
1.Insufficient water level in the sterilization chamber.
2.Water conductivity lower than 10 μs/cm.
3.Dirt or air leakage in the exhaust solenoid valve.
Solutions:
1.Add water into the sterilization chamber.
2.Add a small amount of tap water into the sterilization chamber.
3.Dismantle and inspect the spool of the exhaust solenoid valve (to be performed by aftersales engineers).
- n03
Possible Causes:
1.Invalid system time.
2.Flat battery.
3.Timekeeping battery depleted after longterm instrument inactivity.
Solutions:
1.Reset the system time.
2.Replace the battery.
3.Allow the instrument to stand by for one hour before use after prolonged inactivity.
- n06
Possible Cause: Chamber door not locked upon cycle startup. Solutions: Relock the sterilization chamber door.
- n8
Possible Cause: Water tank is short of water.
Solutions: Place the water inlet pipe on the right side of the instrument into water, then press the “WATER ADDING” key to fill the water tank.
- n09
Possible Cause: The filter element of the biosafetymodel autoclave has reached its service life cycle count.
Solutions: Replace the biosafety filter element.
- n11
Possible Causes:
1.Contaminated waterlevel sensor.
2.Dirty water inside the sterilization chamber.
3.Chamber water has not been replaced for more than two weeks.
Solutions: Clean the waterlevel sensor and sterilization chamber, and drain all water from the chamber.